Fire Fighting Equipments in India, Fire and Safety Products, Fire Safety Products, Safety Products, Safety Equipments, Fire Safety Equipments
Home ::
Links ::

 

Polyon tox
                            
Drager Polytron 7000
Designed in co-operation with our customers the Dräger Polytron 7000 is the leader of the next generation of intrinsically safe toxic and oxygen gas detectors. The goal behind the development was to service our customers and their wide variety of requirements and applications on a single, modular platform. The Dräger Polytron 7000 is able to adapt in line with the changing specifications of the
application.
 
Drager Polytron 3000
Dräger Polytron 3000:
Intrinsically safe gas
detector for toxic gases
and oxygen in ambient air.
 
GAS RANGE
Acid Compounds 0 to 3 / 10 ppm
Ammonia (NH3) 0 to 100 / 300 / 1000 ppm
Boron Trichloride (BCl3) 0 to 10 ppm
Carbon Monoxide (CO) 0 to 100 / 300 / 1000 ppm
Chlorine (Cl2) 0 to 1 / 10 / 25 ppm
Diborane (B2H6) 0 to 0.5 ppm
Hydride (PH3) 0 to 0.3 / 1 / 10 ppm
Hydrogen (H2) 0 to 1000 / 3000 ppm
Hydrogen Sulfide (H2S) 0 to 20 / 50 / 100 ppm
Hydrazine (N2H4) 0 to 1 ppm
Nitrogen Dioxide (NO2) 0 to 10 ppm
Nitrogen Monoxide (NO) 0 to 50 ppm
Organic Vapours 0 to 50 / 100 / 200 ppm
Oxygen (O2) 0 to 5 / 25 / 100 Vol%
Ozone (O3) 0 to 0.5 ppm
Sulfur Dioxide (SO2) 0 to 10 ppm

TECHNICAL INFORMATION

Type Intrinsically safe transmitter for electrochemical sensors
Gases and Ranges Toxic gases and oxygen in various ranges
Output Analog                       -      4 to 20 mA
Maintenance signal    -     3.4 mA constant
Fault signal                 -     < 3.2 mA
Power Supply 12 to 30 VDC, 2-wire
Ambient Conditions Transmitter only, for sensors see separate sensor data sheet
Temperature      -      – 40 to + 65 °C / – 40 to + 150 °F

Pressure           -      700 to 1300 mbar / 20.7 to 38.4 inch Hg
Humidity            -       0 to 100 %RH, non condensing
Enclosure IP 66 / 67, NEMA 4, GRP housing
Size (H x W x D, approx.) 170 x 130 x 130 mm / 6.7” x 5.1” x 5.1”
Weight (approx.) 900 g / 2 lbs.

Approvals
ATEX II 1G EEx ia IIC T6/T4: – 40 to + 40 / + 65 °C
II 3G EEx nL IIC T6/T4: – 25 to + 40 / + 65 °C
IECEx EEx ia IIC T6/T4: – 40 to + 40 / + 65 °C

UL
Class I, Div 1, Group A, B, C, D
Class II, Div 1, Group E, F, G
T6/T4: – 40 to + 40 / + 65 °C
CSA* Class I, Div 1, Group A, B, C, D
Ex ia T6/T4: – 40 to + 40 / + 65 °C

CE mark
electromagnetic compatibility (directive 89/336/EEC)
*pending









The Dräger Polytron 3000 is an intrinsically safe gas detector for the continuous monitoring of toxic gases and oxygen in ambient air. It is the first part of a new generation of gas detectors developed on a modular platform. Communication to the central control system is done via a 4 to 20 mA signal. The Dräger Polytron 3000 is available with or without display, in over 60 versions for different gases and set ranges.

Simple, quick installation The two component concept of a Docking Station and a Dräger Polytron 3000 electronics saves time and money. The Docking Station can be pre-installed – mounting and wiring it
into place separately – while protected by a rain and dust cover until commissioning. At commissioning, the Dräger Polytron 3000 electronics is fixed by a quicklock mechanism into the Docking Station, the sensor inserted - and the system is ready for operation.

Easy to operate The Dräger Polytron 3000 is easy to handle with simple maintenance procedures. The DrägerSensors are equipped with a data memory where the calibration information is stored. This allows the Dräger Polytron 3000 to either accept a precalibrated sensor, or the sensor can be calibrated with the transmitter. Digital temperature compensation of the sensor signal is automatically performed.
 

 
 

 
   © Copyright 2007 2008 Unisafe Services. Email  deshmukh@unisafeservices.com